ERROR CODES
DIAGNOSTICS FOR RT300/RT310 ROTARY TEMPERATURE TRANSMITTER SYSTEMS
POWER STATUS
Stationary and rotary power in spec
Rotary power out
of spec
CORRECTIVE ACTION:
Check rotor/stator spacing coax connections.
Stationary power out of spec
CORRECTIVE ACTION:
Check power source
System not powered
CORRECTIVE ACTION:
Check power source, power connections.
Insufficient rotary power or data not received
CORRECTIVE ACTION:
Check rotor/stator spacing, coax connections.
DATA STATUS
Digital Transmission is error-free
Data not received
CORRECTIVE ACTION:
Check rotor/stator spacing coax connections.
RTD STATUS
No errors detected
Rotary side error:
RTD Channel 1 out of range (including open or shorted)>
CORRECTIVE ACTION:
Check RTD, connections
Open circuit in 4-20mA loop on Channel 1
Check connections and continuity of current loop